Friday, October 8, 2010

Optical temperature measurement in-situ TEM

I promised to report back on the latest ongoing project, i.e. temperature measurement in-situ transmission electron microscopy (TEM).

Background
At our lab we usually do a lot of in-situ TEM experiment with special types of sample holders that have a movable probe. The probe is used to contact and characterize nanomaterials and in these experiments most often electrical current is used to probe and heat up the sample. However measurements of the actual temperature induced by this electrical heating is problematic in these types of two-probe setup. We know the input electrical power and can model and estimate the temperature based on this. But dissipation of heating energy in contact resistances make this a rather uncertain method. A direct temperature method would be most useful and by introducing a optical fiber into the system, this is possible for higher temperatures, i.e. above 1000 Kelvin, by measuring the blackbody spectra.


Benefits of putting an optical fiber inside the TEM
One can think of several benefits:


Method
As mention before, we use special types of sample holders. That is modified versions of the Nanofactory Instruments AB, se figure 4, holders. Because these holders already have electrical wiring to control the probe in front of the holder they are easily modified to fit more wiring or as in this case an optical fiber. This project was driven as a two student projet at bachelor university level in which the students modified the holder with the insertion of an optical fiber during a period of ten weeks. Great thanks to these two good students of the Mid Sweden University.


The most obvious problem is how to get the optical fiber inside the vacuum system of the TEM without introducing a leak. We chose to build our solution on an optical fiber patch cord 200um of VIS/NIR type from Edmund Optics. This saved us time as we did not need to assemble the optical SMA contacts that the spectrometer uses.


Most technical details can be seen in the drawing shown below, figure 1, so here I only write some supporting information. The patch cord is held in place by fitting and squeezing  an O-ring tight around the cord with a screw metal plate clamping, se figure 2. This to avoid unwanted tension in the accutal fiber that goes through the sample holder.




Figure 1 (top), the schematic drawing of the modified sample holder with an optical fiber. Figure 2 (left), The backside of the sample holder, here one can se the screw clamping of the fiber patch cord at the top.
Figure 3 (right), the front of the sample holder, here one can se the moving probe (right) used for contacting and heating nano materials. At the top one can se the optical fiber, here with light coming out in the end of it. In front of the fiber is placed a 45 degree angle rod mirror coated with aluminum this to direct the black body radiation from the sample (center of image, oposite to the probe) into the fiber. Figure 4 (bottom), the sample holder in hole. The front to the left and the modified backside of the holder to the right in image. This holder is used in a JEOL-2000fx transmission electron microscope.

Back of sample holder
The fiber patch cord is unwrapped and the fiber is left exposed at the front of the clamping position. Here the fiber is threaded through a rubber-plug and this plug is fixated in a tapered hole, this gives a high vacuum tight seal, some Fomblin grease was used to make things slide more easily. By the way remember that if you instead have an ultra high vacuum system, rubber is not allowed. After the plug the fiber is continuing towards into the front of the sample holder. To assemble this a fishing line was used and threaded from the oposite direction and glued to the top of the fiber to help threading and guiding the fiber through the holder. 

Front of sample holder
The fiber is threaded through and comes out in the front of the sample holder at the side of the probe, se figure 3 and cut with a tweezer-cutter to proper length. First we thought that some fixation device was needed to hold the fiber in place but since the 200um optical fiber is quit stiff this was not needed. We also did not notice any problems with charging due to the e-beam exposure of the fiber.
To direct the light produced by the heated sample into the fiber a mirror was used. We bought a rod mirror of diameter 1mm from Edmund Optics for this purpose, it is aluminum coated so charging by the e-beam would be less problematic.  We also bought a prism but did not use this for the same reason that the beam might charge this uncoated prism, se figure 5. The mirror was glued with electrical conductive glue onto a Pt/Ir stiff wire that was fitted into a metall tube that was in-turn glued to the side of the holder and at the front of the fiber, this to allow for proper adjustment of the mirror so that the radiation from the sample could be directed into the fiber, se figure 3. Light was led through the fiber in the oposite direction, i.e. from the spectrometer side, to make proper alignment of the fiber-mirror-sample system.
Figure 5, rod mirror with aluminum coating and a diameter of 1mm (left) and the prism that was not used  due to believed electric beam charging (right).

Spectrometer
We used a previously bought spectrometer from Edmund Optics for the first tests, it is a relabeled BWTEK CCD-spectrometer of VIS-NIR type. It has sensitivity up to 1000nm wavelength. The sensitivity drops almost linearly from 500nm to 1000nm by 75% which is of course quit bad for blackbody experiments which has it's peaks around 1000nm, but it did well for a first test result as seen figure 6. 

Results
What we did was to heat up, by driving an electrical current through, a sample of Mo6S3I6 nanowires. The sensitivity was not enough to record the blackbody radiation intensity from a single nanowire bundle but only when severals was heated at the same time. The spectrometer gives a curve that resembles quit well a curve from previously measured well defined temperature source of 1300 Kelvin (furnace) so the conclusion is that the samples was heated to the same temperature, se figure 6.
Figure 6 (left), spectrometer data of the first in-situ TEM experiments when a Mo6S3I6 nanowire sample is heated up to 1300 Kelvin, as proven by the calibration curve from a furnace test. Figure 7 (right) CNT 's heated up to almost 2800 Kelvin, compared to a tungsten lamp source spectra, the lamp spectra is somewhat flooded in the detector counts at its peak.

If you know some blackbody radiation, se wikipedia for example, theory you know that the peak of the radiation spectra would be 1000nm for a tungsten lamp at 2800 Kelvin and the peak in our diagram above was found at a shorter wavelength thus indicating an even higher temperature than this. Of course this is not the case thus the observed peak is due to the spectrometers dropping sensitivity for increasing wavelength, one can correct this curve to a known temperature source so it shows the real blackbody curve, however this is not shown as this beyond the topic of this blog post. We also tried heating up carbon nanotubes, here also with severals at the same time but here we achieved a must higher temperature of almost 2800 Kelvin as seen if one compares the data with a tungsten lamp source at 2800 Kelvin, se figure 7. Due to the higher temperature the blackbody intensity is increased dramatically which is beneficial for the detector. Previously we used the maximum integration time of the spectrometer, that is 60s, but here we could lower this to 10s and yet still, half of the maximum count of the detector was achieved.

Conclusion
The method of using an optical fiber patch cord to introduce in-situ temperature measurements in the TEM worked. Blackbody radiation intensity was recorded for a Mo6S3I6 sample heated to a temperature of 1300 Kelvin as well as a carbon nano tube sample heated to 2800 Kelvin. The used spectrometer was not optimum, a better chice would have been a Prime X spectrometer or a Sol 1.7 spectrometer from BWTEK which has a better sensitivity in the range of 800nm to 1200nm. Higher wavelength sensitivity will probably not be needed as blackbody spectra in these wavelength would be of to weak intensity anyway.

Whats next
Currently we are working on making the hole system of this in-situ TEM probing mobile by utilizing a PXI-measurement system from National Instruments.

Good sailing on the sea of knowledge /Dr. Hummelgård

Wednesday, August 25, 2010

Heating holder ruthenium-salt experiment

In a previous post I reported on how we made an old heating sample holder Jeol EM-SHH to function by replacing the electronic with LabView and PXI measurement system. Here I report on the first results of a new upcoming experiment.

Background
Ruthenium together with different other substances are often used due to it's catalytic properties and is for example applied in electrodes used for electrochemical reactions. In electrochemical processes, large area of the electrode is preferred as this scales with production. To create large areas, usually porous structures on the electrode surface are constructed by different processes, heating in furnace is one of them.


A thin film of the salt-mixture were deposited on a silicon nitride TEM grid. Here we test the heating holder by heating up this ruthenium-salt mixture to above 700 °C.

Result
We made a movie of the experiment, at the bottom of the movie screen you see the heating temperature.
Two important observations were made:

  1. At 14 seconds into the movie, the ruthenium-film crystalizes at the beginning the film was amorphous.
  2. Nanoparticles were formed late in the movie, and at 41 seconds, these particles can be seen to coalescence.  

Conclusion
The sample heating holder appears to function properly.

Supplementary
The metallic springs used to fix the sample in the holder were lost, before I received it. New springs were made by winding a 0.25 mm platinum/iridium wire around a small screwdriver, this seemed to work properly.

Whats next
Next I will repport on how to introduce an optical fiber into the electron microscope, used for temperature measurements or laser treatment in-situ.

Tuesday, June 29, 2010

Jeol EM-SHH (heating holder) fix

A sample heating holder
Is a device that makes it possible to heat the TEM-grid during observation inside the TEM. These special holders can usually go up to temperatures beyond 1000°C with make’s it pretty neat to study anealing, crystalisation and other related phenomena for metals or other types of materials.


-The good’s, I acquried an used heating holder of Jeol fabricate and it fits our microscope, the 2000fx that is.

-The bad’s, no manual, no power supply only the actual holder was devliered.

Solution:
1. Use National Instruments PXI system as replacement for the missing power supply unit.
+. This gives a bonus of complete integration to the rest of our system setup, sync with video recording, logging etc.

Some subproblems to consider:
Find out the cable configuration on the back of the holder. These holder’s usually contains a thermocouple (TC) to measure (two wires) the temperature and a filament (another two wires) to heat the sample with. In figure 2 below one sees a photo of the back of the holder, showing the connected wires from the grey cable that goes to the connector that fits in the power supply unit. An ohm meter will give you which wires belong to each other but to find out which pair is the thermocouple and which is the filament one need to measure the resistance temperature dependency of the pairs. I used the PXI system to measure the temperature dependency. In this case it was the RED and WHITE wires. This means that the remaining pair BLACK and GREEN are used for the filament.

I decided to remove the old contact on the heating holder and replace it with a new one of Fischer type.
Figure 1, the Fischer connector (103 type) that was assembled onto the heating holder cable.
To find out what type of TC is being used
I search the Internet and found a paper describing that a similar holder used a thermocouple of PT-PT-RH(13%) combination, R-type that is.

The paper was:
Crystallization of amorphous thin low pressure chemical vapour deposition silicon films' in situ TEM measurement of grain growth rates
J. P. GUILLEMET, B. DE MAUDUIT, B. PIERAGGI, E. CAMPO and E. SCHEID
JOURNAL OF MATERIALS SCIENCE LETTERS 12 (1993) 910-912

I quick test to see if this is also right for my holder was done. I measured the temperature using the R-type configuration of the thermocouple and melted a small piece of Indium metal that I have placed on the TEM grid. According to the temperature reading of the thermocouple, the metal melted at 155°C, table value states 156.6°C for Indium so this is within the error range and the thermocouple is most likely of R-type.
Figure 2, The back side of the heating holder RED, WHITE thermocouple and BLACK, GRREN heating filament.






The thermocouple circuit
Just to mention som technical details about the temperature measurement. I use a PXI system with the National Instrument's PXI-6229 multifunctional M-series DAQ in combination with a SCB-68 connector block. This is a sleek setup for thermocouple measurements, the LabView software has preconfigured scaling for thermocouples, you just select which type you wanna use, and the SCB-68 connector block has built in cold-junction-compensation by utilizing a LM34 temperature sensor. Because the thermocouple is differential you easily get noise built up in the wires, so I soldered a pair of 10kOhm resistors from the thermocouple lines to ground in the connector box. A low-pass filter with cut of frequency of 4 Hz was also assembled in the connector box. For me this setup gave a signal that is stable to sub ±1°C.

The heating filament circuit
The heating of the filament also need's some signal conditioning setup. First a thought of using a power operational amplifier, but I find out that I already got a ULN2003, a driving stage, in my Racco box so that have to do. This drive-stage has seven channels that works similar to a relay, when a channel is activated on the drive-stage, by a digital TTL signal, it connects the signal to be controlled to ground. I configured all seven channels into a parallel resistors net, where the resistors where configured with the following values: 10, 20, 43, 89, 160, 320 and 650 Ohm. So what does this mean: If I activate the 10 Ohm channel I am sinking an high current due to the low resistance (hot filament) and if I activate a high resistance channel (less hot filament), a smaller current will pass to ground. Because all my seven channels goes through the filament and due to the neat chose of resistors (note the factor of 2x increase between each of them) I can control the heating power level of the filament in 255 step using a binary combination of all the channels.

One thing to consider here is that you limit the heating current by using resistors, the driving stage can sink up to 500mA per channel, so this means that you could get up to 2.5 Watts of heat production in your 10 Ohm resistor, so have this in thought when choosing resistors. I choose a power resistor capable of handling 4 Watts for the 10 Ohm resistor channel, and the 20 Ohm resistor a choose two 1 Watts resistors in series. For the rest the power level is less than 0.5 Watts so for those the power levels are not a problem.

The drive stage is supplied externally with a voltage source.

Figure 3, The SCB-68 connector box with the extra electronics assembled. Top left you se the low-pass filter, (the blue capacitor and resistor) used for the thermocouple, below left is the Fischer connector that goes to the heating holder. The Fischer connector was chosen by me because the one that was on the heating holder was of some old model. To the bottom right you se incoming power supply for the ULN2003, drive stage, the IC located in the middle. Below the IC is the parallell resistor net with a high power resistor to the outer most right of them.
Figure 4, I used Multisim to model the electric circuit to see if the heating powers of the resistors were below the maximum specifications of them.  
The result
I configure a PID control in LabView to control the heating of the filament with the measured  temperature as feed back, a simple program that doesn't need to be shown here I think. This program is now linked to my other LabView program so I will get movie and logging capabilities of a heating experiment.

I tested to heat up the filament, as can bee seen in the graph below, the temperature is considerably higher inside the TEM (vacuum) compared to outside were you will also have a cooling effect by the air.
For air 680mA of heating current gives a temperature of 155°C, were inside the TEM this current level will give you a temperature of roughly 800°C.
Figure 5, Measured temperature as function of heating electric current. Resistance curve is the modeled total resistance including the limiting resistors net. Voltage curve is the modeled applied voltage to the connector box. Heating power is the modeled heating power that the filament produces.

Update
The resistors I used earlier in this experiment were very old and after some running time they could not withstand the heat. So I bought new 2 Watts resistors, as seen in the figure 6.

Figure 6, Resistors, right in picture, were replaced with fresh ones of 2 Watt type, except the 4W power resistor that was still ok. The new coaxial cable seen in this picture, compared with the last, is only used in an another type of experiment.