Showing posts with label in-situ TEM. Show all posts
Showing posts with label in-situ TEM. Show all posts

Friday, March 23, 2012

The new in-situ TEM-SPM 3000 probe control

Background
I promised you in the last post that this time it would be a little bit more interesting. This is a post about the new control system for the in-situ TEM probe that just recently has been developed, figure 1.



Figure 1, (left) the in-situ TEM probe from Nanofactory instruments. (right) The probe and the new control system fits nicely in a transportable suitcase making the hole system mobile.

The new control system hardware
Is a National Instruments PXI-system with a PMA, portable touch screen monitor and keyboard attached, and the system consist of, se figure 2:

  1. An high-end embedded quad-core controller.
  2. Firewire camera adapter (PXI-8252).
  3. Three Source Measurement Unit (SMU) 100 Volts cards (PXI-4132).
  4. One DMM 7.5 digits, high-voltage multimeter card (PXI-4071).
  5. One multiplexer; mechanical relay switch card (PXI-2593).
  6. Interface-cabling.
  7. The Nanofactory probe (that fits a JEOL-2000fx TEM).
Figure 2, The PXI system from National Instruments being used for the probe control, here seen with the cables connected.


Methods

Key ideas behind the solution:
  • The SMU cards supplies the high-voltage required to control the piezo-tube probe.
  • The main purpose of the switch-card is to short-circuit the piezo after its being charged by the SMU's, assuring for the necessary quick physical return movement of the probe that makes the sliding motion function properly.
  • Another advantage of also utilizing a switch-card is that the number of SMU-cards for the probe control can be reduced to a minimum of two.
  • Capacitors are used to preserve charge on the piezo after that the voltage sources has been disconnected. This is possible due two the extreme low current consumption of the piezo.

Why use SMU-cards?
Well, the main advantage for this is that this specific card can supply 100 Volts, so two cards will give you 200 Volts, which is enough for controlling the probe. Other advantages of the SMU-cards is good accuracy and current sourcing capabilities, which could perhaps be used in a future experiment for controlling the piezo by the current-charging method instead of todays voltage method, this reduces hysteresis effect of the piezo. Also the SMU-card is nice to use as a bias-tool for the probe, if one has one extra card over the two already used that is.

The idea behind the switch card
The main idea behind using a switch-card is to disconnect the voltage sources that have been used to charge the piezo-tube scanner and then discharge the piezo over itself, i.e. short-circuit it. Figure 3 shows the first setup tested which worked correctly. However it was soon realized that the grounding post was not needed and two voltage-sources where put in series to increase the supplied voltage up to 200 Volts.

Figure 3, The main idea to use the switch to short-circuit the piezo allowing for a quick discharge.

The SMU cards is connected by the switch cards to the piezo, the piezo is charged by a voltage ramping function, then disconnected followed by discharge over it self, figure 4 shows the oscilloscope view for a complete pulse, voltage ramp and discharge that is.

Figure 4, Oscilloscope view of the pulses created by the SMU and switch cards. Timescale 5ms/square, peak-to-peak 200 Volts. The speed of the pulses can be increase up to a maximum of about 5ms/pulse.

Nothing else can be made to discharge the piezo faster than this short-circuiting method, previously the "discharge" was made by dropping signal level on the source to zero as fast as possible, but this always gave slow discharge rates, fatal for the sliding motion, in which the probe will just vibrate and no sliding will occur. Some positive effect on the discharge speed was seen by manipulating the ramping function and switching from saw-tooth style to a more exponential looking style of the voltage-function, why this gives better results might be due to that more aggressive pulses easier travels through the often built in filters thats included in comercial hardware, but this is just a guess and has not been further investigated. For the new discharge method, ramp-function style seems to be of no importans.

Two SMU-cards to drive five piezo channels?
If a switch card is being utilized for the discharging of the piezo, it can also be used to reduce the number of voltage sources used for the piezo. Normally one would use one source for each channel, that is: two for x-direction, two for y-direction, and one for z-direction.

If you can store the charge on the piezo by utilizing capacitors you can then disconnect the voltage source from one channel and then switch to connect the source to another. Using capacitors is possible because the self discharging leakage current of the piezo is minimum especially if you run the probe in vacuum as you normally do. a few hundred nanoFarads would be enough to guarantee the preservation of charge of the piezo. (internal resistance in GigaOhm range when used in air) Figure 5 shows the connection of the capacitors to the piezo and the switch-card. Interestingly, by using this star-capacitor configuration one seems to get a more similar connection for the Z channel compared to the X and Y, thus with the capacitors one get what I called the missing Z- channel, so now there will be six instead of five channels: X+/-, Y+/-, and the new Z+/Z- channels. This simplifies connection and charging of piezo.

Figure 5, The star-configuration of the capacitors connected to the five piezo-channels and the switch-card. Switches shows schematically how to switch between X,Y, and Z movement direction of the probe. Now the Z channel consist of two connectors Z+ and Z-.

If you study figure 5 in detail you will also notice that for the Z direction mode both SMU-sources can be used by reversing the direction of one of the sources and putting them in series with each other, giving a total of 200 Volts on the Z-channel, compared to the default which is one source and 100 Volts.

Figure 6, The connectors that connects the probe to the PXI-system. (left) two red and blue connectors for the SMU-cards, a bunch of MCX contacts that connects to the switch (not seen), the Fisher contact that connects to the probe (silver). (right) The red connector disassembled showing that two of the capacitors has been hidden inside the connector, the other three capacitors are hidden in the blue connector this to assures a nice look of the cabling interface. Later I decided to make a change of the cabling by replacing the gray signal cable that goes to the probe with a bunch of  individual RG174 cables for each signal, this improved signal/noise level somewhat.


Figure 7, shows how to setup the internal relays in the multiplexer card depending on if one wants to run X,Y and Z mode. When running the software this will be transparant to the user and depending on which control the user uses the multiplexer card selects and jumps between proper configuration automatically. If you look att the schematics you see that one side or half the multiplexer is free for use, so controlling of other equipment or perhaps a double probe is a possible extension. The lifetime of the relays has been considered and according to the spec. 300.000/5.000.000 electrical/mechanical cycles are estimated lifetime so they should last for quit a long time. For information, now after that the development is completed the relay switching consumption has been up to around 50.000 cycles/relay and then the intertial sliding function was used all the time for testing in one months development time. 



Figure 7, schematics of the internal configuration of the switch card for the Z,X, and Y mode control of the probe. (click to enlarge)


The discharging current when short-circuiting the piezo has also been considered. By introducing extra capacitors, as done above, you introduce extra charge that has to be taken care of during the discharge of the piezo. The piezo which has a capacitance of around 1nF can therefore be neglected in contributing to the discharge current: if one uses two 100nF capacitors, neglecting the charge of the piezo and estimating the discharge rate to be 0.1ms at 200 Volts the short-circuit current level will be 400mA (mean value), the relays are constructed for 500mA breaking-current so it will probably be okay. If this will be a problem in the future, a current limiting resistor can be put in series with the relays, limiting the discharge current, but this will of course be at the expense of increase of discharge time. Better is to reduce the value of the capacitances, now I uses two 150nF capacitors, but according to test, se figure 8, lower values will also do. Inner resistance of the piezo is GigaOhm range so if one estimates the leakage current to be say 100pA then a 200 Volts charge over two 20nF capacitors will discharge by 1% over a time of 13 minutes and in that case the short-circuit current will be 80mA for the same short-circuit time period as stated above.

Results

Check out the movie on this subject at YouTube channel. The movie shows the first basic test of the coarse and fine movement of the probe in an few in-situ TEM experiment.

Speed of probe in coarse motion (inertial sliding) is about 30-120 um/s.

Minimum voltage (across the piezo) required for successful inertial sliding operational:
Mode   Potential
----------------------
X+ 124 volts
X- 134 volts
Y+ 138 volts
Y- 158 volts
Z+ 84  volts
Z- bad function*
-----------------------
*, bad function in Z- direction to to wrong tension of the spider-leg springs. Springs were too soft during this test. Later test showed that good motion is achieved at voltage-levels from 140 volts and higher and that sliding still works down to 100 volts, this for all piezo channels.


Discharge rate measurement of the piezo-probe by monitoring the probe position in TEM over time with different electrical configuration: constants voltage applied which shows the piezo hysteresis drift effect, zero nF (no capacitors used) the piezo discharge itself and starts to move, 1nF, 47nF to 470nF values of capacitances showing that the higher values is enough to keep the piezo from discharging over a time period of 10 minutes, figure 8.
Figure 8, discharge rate of piezo tested for different electrical configurations. 47nF seems to be enough to prevent discharge of the piezo to that level that it will result in probe drift during the measurement time.

Conclusion
The new TEM-SPM 3000 control system works and will now be used in future experiments. If monitoring the probe in TEM moving artefacts (drifts or jumps) can not be seen when switching between X,Y, and Z mode channels of the piezo. The probe also seems stable without drift for long time periods, except for piezo hysteresis that is. Later I will perhaps make a post about how the LabView software. 

Wednesday, August 25, 2010

Heating holder ruthenium-salt experiment

In a previous post I reported on how we made an old heating sample holder Jeol EM-SHH to function by replacing the electronic with LabView and PXI measurement system. Here I report on the first results of a new upcoming experiment.

Background
Ruthenium together with different other substances are often used due to it's catalytic properties and is for example applied in electrodes used for electrochemical reactions. In electrochemical processes, large area of the electrode is preferred as this scales with production. To create large areas, usually porous structures on the electrode surface are constructed by different processes, heating in furnace is one of them.


A thin film of the salt-mixture were deposited on a silicon nitride TEM grid. Here we test the heating holder by heating up this ruthenium-salt mixture to above 700 °C.

Result
We made a movie of the experiment, at the bottom of the movie screen you see the heating temperature.
Two important observations were made:

  1. At 14 seconds into the movie, the ruthenium-film crystalizes at the beginning the film was amorphous.
  2. Nanoparticles were formed late in the movie, and at 41 seconds, these particles can be seen to coalescence.  

Conclusion
The sample heating holder appears to function properly.

Supplementary
The metallic springs used to fix the sample in the holder were lost, before I received it. New springs were made by winding a 0.25 mm platinum/iridium wire around a small screwdriver, this seemed to work properly.

Whats next
Next I will repport on how to introduce an optical fiber into the electron microscope, used for temperature measurements or laser treatment in-situ.

Tuesday, June 29, 2010

Jeol EM-SHH (heating holder) fix

A sample heating holder
Is a device that makes it possible to heat the TEM-grid during observation inside the TEM. These special holders can usually go up to temperatures beyond 1000°C with make’s it pretty neat to study anealing, crystalisation and other related phenomena for metals or other types of materials.


-The good’s, I acquried an used heating holder of Jeol fabricate and it fits our microscope, the 2000fx that is.

-The bad’s, no manual, no power supply only the actual holder was devliered.

Solution:
1. Use National Instruments PXI system as replacement for the missing power supply unit.
+. This gives a bonus of complete integration to the rest of our system setup, sync with video recording, logging etc.

Some subproblems to consider:
Find out the cable configuration on the back of the holder. These holder’s usually contains a thermocouple (TC) to measure (two wires) the temperature and a filament (another two wires) to heat the sample with. In figure 2 below one sees a photo of the back of the holder, showing the connected wires from the grey cable that goes to the connector that fits in the power supply unit. An ohm meter will give you which wires belong to each other but to find out which pair is the thermocouple and which is the filament one need to measure the resistance temperature dependency of the pairs. I used the PXI system to measure the temperature dependency. In this case it was the RED and WHITE wires. This means that the remaining pair BLACK and GREEN are used for the filament.

I decided to remove the old contact on the heating holder and replace it with a new one of Fischer type.
Figure 1, the Fischer connector (103 type) that was assembled onto the heating holder cable.
To find out what type of TC is being used
I search the Internet and found a paper describing that a similar holder used a thermocouple of PT-PT-RH(13%) combination, R-type that is.

The paper was:
Crystallization of amorphous thin low pressure chemical vapour deposition silicon films' in situ TEM measurement of grain growth rates
J. P. GUILLEMET, B. DE MAUDUIT, B. PIERAGGI, E. CAMPO and E. SCHEID
JOURNAL OF MATERIALS SCIENCE LETTERS 12 (1993) 910-912

I quick test to see if this is also right for my holder was done. I measured the temperature using the R-type configuration of the thermocouple and melted a small piece of Indium metal that I have placed on the TEM grid. According to the temperature reading of the thermocouple, the metal melted at 155°C, table value states 156.6°C for Indium so this is within the error range and the thermocouple is most likely of R-type.
Figure 2, The back side of the heating holder RED, WHITE thermocouple and BLACK, GRREN heating filament.






The thermocouple circuit
Just to mention som technical details about the temperature measurement. I use a PXI system with the National Instrument's PXI-6229 multifunctional M-series DAQ in combination with a SCB-68 connector block. This is a sleek setup for thermocouple measurements, the LabView software has preconfigured scaling for thermocouples, you just select which type you wanna use, and the SCB-68 connector block has built in cold-junction-compensation by utilizing a LM34 temperature sensor. Because the thermocouple is differential you easily get noise built up in the wires, so I soldered a pair of 10kOhm resistors from the thermocouple lines to ground in the connector box. A low-pass filter with cut of frequency of 4 Hz was also assembled in the connector box. For me this setup gave a signal that is stable to sub ±1°C.

The heating filament circuit
The heating of the filament also need's some signal conditioning setup. First a thought of using a power operational amplifier, but I find out that I already got a ULN2003, a driving stage, in my Racco box so that have to do. This drive-stage has seven channels that works similar to a relay, when a channel is activated on the drive-stage, by a digital TTL signal, it connects the signal to be controlled to ground. I configured all seven channels into a parallel resistors net, where the resistors where configured with the following values: 10, 20, 43, 89, 160, 320 and 650 Ohm. So what does this mean: If I activate the 10 Ohm channel I am sinking an high current due to the low resistance (hot filament) and if I activate a high resistance channel (less hot filament), a smaller current will pass to ground. Because all my seven channels goes through the filament and due to the neat chose of resistors (note the factor of 2x increase between each of them) I can control the heating power level of the filament in 255 step using a binary combination of all the channels.

One thing to consider here is that you limit the heating current by using resistors, the driving stage can sink up to 500mA per channel, so this means that you could get up to 2.5 Watts of heat production in your 10 Ohm resistor, so have this in thought when choosing resistors. I choose a power resistor capable of handling 4 Watts for the 10 Ohm resistor channel, and the 20 Ohm resistor a choose two 1 Watts resistors in series. For the rest the power level is less than 0.5 Watts so for those the power levels are not a problem.

The drive stage is supplied externally with a voltage source.

Figure 3, The SCB-68 connector box with the extra electronics assembled. Top left you se the low-pass filter, (the blue capacitor and resistor) used for the thermocouple, below left is the Fischer connector that goes to the heating holder. The Fischer connector was chosen by me because the one that was on the heating holder was of some old model. To the bottom right you se incoming power supply for the ULN2003, drive stage, the IC located in the middle. Below the IC is the parallell resistor net with a high power resistor to the outer most right of them.
Figure 4, I used Multisim to model the electric circuit to see if the heating powers of the resistors were below the maximum specifications of them.  
The result
I configure a PID control in LabView to control the heating of the filament with the measured  temperature as feed back, a simple program that doesn't need to be shown here I think. This program is now linked to my other LabView program so I will get movie and logging capabilities of a heating experiment.

I tested to heat up the filament, as can bee seen in the graph below, the temperature is considerably higher inside the TEM (vacuum) compared to outside were you will also have a cooling effect by the air.
For air 680mA of heating current gives a temperature of 155°C, were inside the TEM this current level will give you a temperature of roughly 800°C.
Figure 5, Measured temperature as function of heating electric current. Resistance curve is the modeled total resistance including the limiting resistors net. Voltage curve is the modeled applied voltage to the connector box. Heating power is the modeled heating power that the filament produces.

Update
The resistors I used earlier in this experiment were very old and after some running time they could not withstand the heat. So I bought new 2 Watts resistors, as seen in the figure 6.

Figure 6, Resistors, right in picture, were replaced with fresh ones of 2 Watt type, except the 4W power resistor that was still ok. The new coaxial cable seen in this picture, compared with the last, is only used in an another type of experiment.

Tuesday, May 4, 2010

Tuning fork force measurement

As mentioned in a previous post, we want to include a quartz tuning fork to be able to directly measure force in-situ TEM. From the easy PLL system that controls the tuning fork a output signal proportional to the frequency shift that occurs due to an mechanical interaction force applied to the tuning force.
According to the PLL manual the frequency shift dF is given by:


dF=K*U
(1)

were U is the detector output signal voltage on the PLL system. Here was K=12.9°/V. If the measured spring Kts constant is low compared to the spring constant Kquartz of the tuning fork, shift in the measured spring constant over one cycle of the amplitude on the tuning fork can be neglected and if so the measured spring constant due to the applied force is given by:

Kts=2*Kquartz*dF/F0
(2)

Were F0 is the resonance frequency of the tuning fork. The result from contacting the tuning fork with the TEM-SPM probe is shown in the plot below were the measured spring constant (of the applied force) is plotted as function of the probe movement. Red and blue curves corresponds to probe moved in forward direction, i.e against the tuning fork, and in reverese direction, i.e away from the tuning fork.

If one integrate (sum) the measurement points one will get the applied force as function of probe movement, as seen in plot below.


One can notice that the forward and reverse curves do not follow the same path in the close-contact region, i.e to the right in plots.

Wednesday, May 13, 2009

My thesis is finished

My thesis "in-situ TEM probing of nanomaterials" is now finished. In Sweden, a thesis is a comprehensive summary of the included scientific work. The paper versions of the thesis consists of two parts the summary, part one, with the scientific papers, part two, attached into one volume. Here is a link to the comprehensive summary part of the thesis.