Thursday, July 15, 2010

Digital images from old SEM/STEM detectors and CRT's

Background
The TEM that we have, the Jeol 2000fx is equipped with two extra detectors, one secondary electron detector, i.e. SEM imaging, as well as a STEM detector at the bottom of the microscope.
Because the system are from the 80's the imaging from these detectors is (read "was") accomplished by using photographic film to copy the image shown on the CRT. An extra CRT monitor specifically used for this task is located in a free standing box in which the camera with light sensitiv polaroid film is attached to.

Nowadays, everything is digital, and polaroid imaging is therefore not very user friendly or timesaving. Besides, one of the two CRT's used for setting upp the image to be captured is broke in our system and the other one is almost burned out... -Time to move into the next century of technology.

Electronics
By different means I found out that the ASID panel, i.e the one that controls the scanning of the electron beam during SEM/STEM imaging, has a extra contact hidden behind the backpanel, labeled SG6, figure 1.
Figure 1, After opening the back-panel of the ASID unit the signal generator card (SG) is somewhat exposed as well as the extra contact SG6, used for external scan control.

If one looks at the circuit diagram, figure 2, one can se that the external scan mode is activated by connecting pin 5 of the SG6 to ground. This mode bypass the scan generator function of the ASID panel and allows you to insert your own scan signals at pin 1 and 2 for horisontal and vertical. The signal level happens to be ±10 Volts and this is lucky thus the PXI-6229 card that I use for this task also has this amplitude range limit. To connect to the SG6 I assembled a signal cable from SG6 internally to the front of the ASID, a switch was used to connect pin 5 to ground, and the horisontal and vertical scan signals got two SMB panel connectors, se figure 3.

Figure 2, the circuit diagram, the essential connector pins for external scan mode is marked with red. Click image to enlarge it.

figure 3 There was one free slot on the ASID front-panel that I used for the switch and connectors. (left) the backside of the ready made front-panel and the frontside of it (right).

The detectors each have BNC connector output I just simply detached the regular cables to these ones and connected my own that leads to the same DAQ-card as mentioned above. This electronic setup will results in that the ASID unit will just handle signal conditioning between my DAQ-card and the beam-scan coils, the two (three) CRT's will no longer be needed. The only buttons on the ASID panel in use will be stigmatism correction and detector contrast setting.

LabView
What you want to do is to get LabView to generate two signals, two sawtooth for horisontal and vertical synch signals. I configured this as a single DAQ-task where both are written at the same time to avoid conflicts on the hardware.  During signal generation you also need to collect the detector data, this I configured as a parallel task to one the above. You have one options here and that's to read data faster than you write, this will improve signal noise ratio for you and I made this option as a user "scan speed" selection in the program. I always read data as fast as possible, 200kHz, and depending on what slower scan speed setting one choose, more samples are read for each position on the sample.

You also have another option here and that is to scan one horisontal line then collect the data from the detector or to read the hole image in one sampling. The former gives you ability to handle bigger image sizes, i.e. beyond 512x512 pixels images, because you only need to have one scan line in the DAQ-card buffer each time. The later alternative is of-course much faster. I choose to implement booth, the one-line scan mode for final imaging and the hole-image scan mode for fast preview as this gives you almost 2 frames/s in 128x128 pixels.

As a final option for improving signal to noise ratio I also made integration over several imaging, i.e. adding images together, as an user option.

Results
To test out the new system configuration I inserted a sample of gold-nanoparticles roughly 50nm in diameters, located on a standard TEM grid.
Figure 4, TEM image of gold-nanoparticles (left) and STEM image of the same, but rotated 90° CCV (right).

Figure 5, SEM image of the same sample and region as shown in figure 4, but at slightly lower magnification this time. Resolution is in this quick check seems to be around 50 nm for the SEM imaging mode, but I expect this to improve if one tveek the system to more optimal settings. (the green color is for those of us who misses the green CRT monitor.)
Future work
Due to the in-situ probing that we usually do in our setup we can measure the impact beam current of the sample during SEM scan and make mapping of this as well this is yet to be done and I report on this later.

Tuesday, June 29, 2010

Jeol EM-SHH (heating holder) fix

A sample heating holder
Is a device that makes it possible to heat the TEM-grid during observation inside the TEM. These special holders can usually go up to temperatures beyond 1000°C with make’s it pretty neat to study anealing, crystalisation and other related phenomena for metals or other types of materials.


-The good’s, I acquried an used heating holder of Jeol fabricate and it fits our microscope, the 2000fx that is.

-The bad’s, no manual, no power supply only the actual holder was devliered.

Solution:
1. Use National Instruments PXI system as replacement for the missing power supply unit.
+. This gives a bonus of complete integration to the rest of our system setup, sync with video recording, logging etc.

Some subproblems to consider:
Find out the cable configuration on the back of the holder. These holder’s usually contains a thermocouple (TC) to measure (two wires) the temperature and a filament (another two wires) to heat the sample with. In figure 2 below one sees a photo of the back of the holder, showing the connected wires from the grey cable that goes to the connector that fits in the power supply unit. An ohm meter will give you which wires belong to each other but to find out which pair is the thermocouple and which is the filament one need to measure the resistance temperature dependency of the pairs. I used the PXI system to measure the temperature dependency. In this case it was the RED and WHITE wires. This means that the remaining pair BLACK and GREEN are used for the filament.

I decided to remove the old contact on the heating holder and replace it with a new one of Fischer type.
Figure 1, the Fischer connector (103 type) that was assembled onto the heating holder cable.
To find out what type of TC is being used
I search the Internet and found a paper describing that a similar holder used a thermocouple of PT-PT-RH(13%) combination, R-type that is.

The paper was:
Crystallization of amorphous thin low pressure chemical vapour deposition silicon films' in situ TEM measurement of grain growth rates
J. P. GUILLEMET, B. DE MAUDUIT, B. PIERAGGI, E. CAMPO and E. SCHEID
JOURNAL OF MATERIALS SCIENCE LETTERS 12 (1993) 910-912

I quick test to see if this is also right for my holder was done. I measured the temperature using the R-type configuration of the thermocouple and melted a small piece of Indium metal that I have placed on the TEM grid. According to the temperature reading of the thermocouple, the metal melted at 155°C, table value states 156.6°C for Indium so this is within the error range and the thermocouple is most likely of R-type.
Figure 2, The back side of the heating holder RED, WHITE thermocouple and BLACK, GRREN heating filament.






The thermocouple circuit
Just to mention som technical details about the temperature measurement. I use a PXI system with the National Instrument's PXI-6229 multifunctional M-series DAQ in combination with a SCB-68 connector block. This is a sleek setup for thermocouple measurements, the LabView software has preconfigured scaling for thermocouples, you just select which type you wanna use, and the SCB-68 connector block has built in cold-junction-compensation by utilizing a LM34 temperature sensor. Because the thermocouple is differential you easily get noise built up in the wires, so I soldered a pair of 10kOhm resistors from the thermocouple lines to ground in the connector box. A low-pass filter with cut of frequency of 4 Hz was also assembled in the connector box. For me this setup gave a signal that is stable to sub ±1°C.

The heating filament circuit
The heating of the filament also need's some signal conditioning setup. First a thought of using a power operational amplifier, but I find out that I already got a ULN2003, a driving stage, in my Racco box so that have to do. This drive-stage has seven channels that works similar to a relay, when a channel is activated on the drive-stage, by a digital TTL signal, it connects the signal to be controlled to ground. I configured all seven channels into a parallel resistors net, where the resistors where configured with the following values: 10, 20, 43, 89, 160, 320 and 650 Ohm. So what does this mean: If I activate the 10 Ohm channel I am sinking an high current due to the low resistance (hot filament) and if I activate a high resistance channel (less hot filament), a smaller current will pass to ground. Because all my seven channels goes through the filament and due to the neat chose of resistors (note the factor of 2x increase between each of them) I can control the heating power level of the filament in 255 step using a binary combination of all the channels.

One thing to consider here is that you limit the heating current by using resistors, the driving stage can sink up to 500mA per channel, so this means that you could get up to 2.5 Watts of heat production in your 10 Ohm resistor, so have this in thought when choosing resistors. I choose a power resistor capable of handling 4 Watts for the 10 Ohm resistor channel, and the 20 Ohm resistor a choose two 1 Watts resistors in series. For the rest the power level is less than 0.5 Watts so for those the power levels are not a problem.

The drive stage is supplied externally with a voltage source.

Figure 3, The SCB-68 connector box with the extra electronics assembled. Top left you se the low-pass filter, (the blue capacitor and resistor) used for the thermocouple, below left is the Fischer connector that goes to the heating holder. The Fischer connector was chosen by me because the one that was on the heating holder was of some old model. To the bottom right you se incoming power supply for the ULN2003, drive stage, the IC located in the middle. Below the IC is the parallell resistor net with a high power resistor to the outer most right of them.
Figure 4, I used Multisim to model the electric circuit to see if the heating powers of the resistors were below the maximum specifications of them.  
The result
I configure a PID control in LabView to control the heating of the filament with the measured  temperature as feed back, a simple program that doesn't need to be shown here I think. This program is now linked to my other LabView program so I will get movie and logging capabilities of a heating experiment.

I tested to heat up the filament, as can bee seen in the graph below, the temperature is considerably higher inside the TEM (vacuum) compared to outside were you will also have a cooling effect by the air.
For air 680mA of heating current gives a temperature of 155°C, were inside the TEM this current level will give you a temperature of roughly 800°C.
Figure 5, Measured temperature as function of heating electric current. Resistance curve is the modeled total resistance including the limiting resistors net. Voltage curve is the modeled applied voltage to the connector box. Heating power is the modeled heating power that the filament produces.

Update
The resistors I used earlier in this experiment were very old and after some running time they could not withstand the heat. So I bought new 2 Watts resistors, as seen in the figure 6.

Figure 6, Resistors, right in picture, were replaced with fresh ones of 2 Watt type, except the 4W power resistor that was still ok. The new coaxial cable seen in this picture, compared with the last, is only used in an another type of experiment.

Wednesday, February 11, 2009

Improvment on meassured variables

New way of meassure

By measuring the best known variables, applied bias voltage U, electric current I and the known series resistor with resistance Rs. The unknown variables of the sample, potential drop Up, sample resistance Rs, power in sample Pp was calculated with lesser noise.
$latex U_s=IR_s$
$latex U_p=U-U_s$
$latex R_p=\frac{U}{I}-R_s$
$latex P_p=UI-I^2R_s$
I have remade the plots with this new approach and it clearly shows much better noise-ratio than before. The calulations before were made by using the noisy-value of the measured potential drop over sample.
20090119-02-overview

The sintering image folder at flicker shows you the other plots that were remade.